Mouse Cell Signalling ELISA Kits 6
Mouse Histamine (HIS) ELISA Kit
- SKU:
- MOEB2524
- Product Type:
- ELISA Kit
- Size:
- 96 Assays
- Range:
- 7.8-500 ng/mL
- ELISA Type:
- Competitive
- Reactivity:
- General
Description
Mouse Histamine ELISA Kit
Histamine is a biogenic amine and signaling molecule involved in various physiological processes, including allergic responses, inflammation, and gastric acid secretion. It acts by binding to specific histamine receptors, triggering specific cellular responses. The Assay Genie Mouse histamine ELISA kit allows for the quantitative measurement of histamine levels in mouse samples, such as serum, plasma, tissue lysates, and cell culture supernatants. It provides researchers with a sensitive and reliable method to assess histamine concentrations, enabling the investigation of histamine-related processes and their role in various physiological and pathological conditions in mouse models.
Key Features
Save Time | Pre-coated 96 well plate | |
Quick Start | Kit includes all necessary reagents | |
Publication Ready | Reproducible and reliable results |
Overview
Product Name: | Mouse Histamine (HIS) ELISA Kit |
Product Code: | MOEB2524 |
Size: | 96 Assays |
Alias: | Histamine, HIS |
Detection Method: | Competitive |
Reactivity: | General |
Range: | 7.8-500 ng/mL |
Storage: | 4°C for 6 months |
Note: | For Research Use Only |
Kit Components
Component | Quantity | Storage |
ELISA Microplate (Dismountable) | 8x12 strips | 4°C for 6 months |
Lyophilized Standard | 2 | 4°C/ -20°C |
Sample/Standard Dlution Buffer | 20ml | 4°C |
Biotin-labeled Antibody (Concentrated) | 120ul | 4°C (Protection from light) |
Antibody Dilution Buffer | 10ml | 4°C |
HRP-Streptavidin Conjugate (SABC) | 120ul | 4°C (Protect from light) |
SABC Dilution Buffer | 10ml | 4°C |
TMB Substrate | 10ml | 4°C (Protection from light) |
Stop Solution | 10ml | 4°C |
Wash Buffer (25X) | 30ml | 4°C |
Plate Sealer | 5 | - |
Other materials required:
- Microplate reader with 450 nm wavelength filter
- Multichannel Pipette, Pipette, microcentrifuge tubes and disposable pipette tips
- Incubator
- Deionized or distilled water
- Absorbent paper
- Buffer resevoir
Protocol
*Note: Protocols are specific to each batch/lot. For the correct instructions please follow the protocol included in your kit.
Equilibrate the TMB substrate for at least 30 min at 37°C beforeuse. When diluting samples and reagents, they must be mixed completely andevenly. It is recommended to plot a standard curve for each test.
Step | Procedure |
1. | Add 50µL of Standard, Blank, or Sample per well. The blank well is added with Sample diluent. Solutions are added to the bottom of micro ELISA plate well, avoid inside wall touching and foaming as possible. |
2. | Immediately add 50µL of Detection Reagent A working solution to each well. Cover with the Plate sealer. Gently tap the plate to ensure thorough mixing. Incubate for 1 hour at 37°C. Note: if Detection Reagent A appears cloudy warm to room temperature until solution is uniform. |
3. | Aspirate each well and wash, repeating the process three times. Wash by filling each well with Wash Buffer (approximately 400µL) (a squirt bottle, multi-channel pipette, manifold dispenser or automated washer are needed) and let it sit in the well for 1-2 minutes. Complete removal of liquid at each step is essential. After the last wash, completely remove remaining Wash Buffer by aspirating or decanting. Invert the plate and pat it against thick clean absorbent paper. |
4. | Add 100µL of Detection Reagent B working solution to each well. Cover with the Plate sealer. Incubate for 45 minutes at 37°C. |
5. | Repeat the wash process for five times as conducted in step 3. |
6. | Add 90µL of Substrate Solution to each well. Cover with a new Plate sealer and incubate for 10-20 minutes at 37°C. Protect the plate from light. The reaction time can be shortened or extended according to the actual color change, but this should not exceed more than 30 minutes. When apparent gradient appears in standard wells, user should terminate the reaction. |
7. | Add 50µL of Stop Solution to each well. If color change does not appear uniform, gently tap the plate to ensure thorough mixing. |
8. | Determine the optical density (OD value) of each well at once, using amicro-plate reader set to 450 nm. User should open the micro-plate reader in advance, preheat the instrument, and set the testing parameters. |
9. | After the experiment, store all reagents according to the specified storage temperature respectively until their expiry. |
Sample Preparation
When carrying out an ELISA assay it is important to prepare your samples in order to achieve the best possible results. Below we have a list of procedures for the preparation of samples for different sample types.
Sample Type | Protocol |
Serum | If using serum separator tubes, allow samples to clot for 30 minutes at room temperature. Centrifuge for 10 minutes at 1,000x g. Collect the serum fraction and assay promptly or aliquot and store the samples at -80°C. A |