Human Metabolism ELISA Kits
Human THRSP (Thyroid Hormone Responsive) CLIA Kit (HUES00798)
- SKU:
- HUES00798
- Product Type:
- ELISA Kit
- ELISA Type:
- CLIA Kit
- Size:
- 96 Assays
- Sensitivity:
- 18.75pg/mL
- Range:
- 31.25-2000pg/mL
- ELISA Type:
- Sandwich
- Reactivity:
- Human
- Sample Type:
- Serum, plasma and other biological fluids
- Research Area:
- Metabolism
Description
Assay type: | Sandwich |
Format: | 96T |
Assay time: | 4.5h |
Reactivity: | Human |
Detection method: | Chemiluminescence |
Detection range: | 31.25-2000 pg/mL |
Sensitivity: | 18.75 pg/mL |
Sample volume: | 100µL |
Sample type: | Serum, plasma and other biological fluids |
Repeatability: | CV < 15% |
Specificity: | This kit recognizes Human THRSP in samples. No significant cross-reactivity or interference between Human THRSP and analogues was observed. |
This kit uses Sandwich-CLIA as the method. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human THRSP. Standards or samples are added to the appropriate micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human THRSP and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well successively and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human THRSP, biotinylated detection antibody and Avidin-HRP conjugate will appear fluorescence. The Relative light unit (RLU) value is measured spectrophotometrically by the Chemiluminescence immunoassay analyzer. The RLU value is positively associated with the concentration of Human THRSP. The concentration of Human THRSP in the samples can be calculated by comparing the RLU of the samples to the standard curve.
UniProt Protein Function: | THRSP: Plays a role in the regulation of lipogenesis, especially in lactating mammary gland. Important for the biosynthesis of triglycerides with medium-length fatty acid chains. May modulate lipogenesis by interacting with MID1IP1 and preventing its interaction with ACACA. May function as transcriptional coactivator. May modulate the transcription factor activity of THRB. Belongs to the SPOT14 family. |
UniProt Protein Details: | Protein type:Unknown function Chromosomal Location of Human Ortholog: 11q14. 1 Cellular Component: nucleoplasm; cytosol Molecular Function:protein homodimerization activity Biological Process: transcription, DNA-dependent; regulation of transcription, DNA-dependent; lipid metabolic process; regulation of lipid biosynthetic process |
NCBI Summary: | The protein encoded by this gene is similar to the gene product of S14, a rat gene whose expression is limited to liver and adipose tissue and is controlled by nutritional and hormonal factors. This gene has been shown to be expressed in liver and adipocytes, particularly in lipomatous modules. It is also found to be expressed in lipogenic breast cancers, which suggests a role in controlling tumor lipid metabolism. [provided by RefSeq, Jul 2008] |
UniProt Code: | Q92748 |
NCBI GenInfo Identifier: | 2842710 |
NCBI Gene ID: | 7069 |
NCBI Accession: | Q92748. 1 |
UniProt Secondary Accession: | Q92748,B2R4W7, |
UniProt Related Accession: | Q92748 |
Molecular Weight: | 146 |
NCBI Full Name: | Thyroid hormone-inducible hepatic protein |
NCBI Synonym Full Names: | thyroid hormone responsive |
NCBI Official Symbol: | THRSP |
NCBI Official Synonym Symbols: | S14; Lpgp; THRP; LPGP1; SPOT14 |
NCBI Protein Information: | thyroid hormone-inducible hepatic protein; SPOT14 homolog; spot 14 protein; lipogenic protein 1; thyroid hormone responsive SPOT14; thyroid hormone responsive (SPOT14 homolog, rat) |
UniProt Protein Name: | Thyroid hormone-inducible hepatic protein |
UniProt Synonym Protein Names: | Spot 14 protein; S14; SPOT14 |
Protein Family: | Thyroid hormone-inducible hepatic protein |
UniProt Gene Name: | THRSP |
UniProt Entry Name: | THRSP_HUMAN |
As the RLU values of the standard curve may vary according to the conditions of the actual assay performance (e. g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration (pg/mL) | RLU | Average | Corrected |
2000 | 51872 59570 | 55721 | 55693 |
1000 | 21959 25989 | 23974 | 23946 |
500 | 11764 10300 | 11032 | 11004 |
250 | 5029 5559 | 5294 | 5266 |
125 | 2609 2607 | 2608 | 2580 |
62.5 | 1341 1281 | 1311 | 1283 |
31.25 | 647 701 | 674 | 646 |
0 | 27 29 | 28 | -- |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Human THRSP were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Human THRSP were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 100.72 | 269.70 | 860.22 | 97.91 | 282.59 | 789.86 |
Standard deviation | 9.40 | 22.25 | 60.39 | 8.20 | 27.78 | 63.50 |
C V (%) | 9.33 | 8.25 | 7.02 | 8.38 | 9.83 | 8.04 |
Recovery
The recovery of Human THRSP spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
Serum (n=5) | 83-96 | 90 |
EDTA plasma (n=5) | 98-110 | 104 |
Cell culture media (n=5) | 92-104 | 99 |
Linearity
Samples were spiked with high concentrations of Human THRSP and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
1:2 | Range (%) | 87-101 | 101-117 | 86-98 |
Average (%) | 95 | 109 | 93 | |
1:4 | Range (%) | 86-96 | 92-105 | 84-97 |
Average (%) | 91 | 98 | 90 | |
1:8 | Range (%) | 85-100 | 88-101 | 94-105 |
Average (%) | 92 | 94 | 100 | |
1:16 | Range (%) | 100-110 | 87-101 | 93-106 |
Average (%) | 105 | 93 | 99 |
An unopened kit can be stored at 4°C for 1 month. If the kit is not used within 1 month, store the items separately according to the following conditions once the kit is received.
Item | Specifications | Storage |
Micro CLIA Plate(Dismountable) | 8 wells ×12 strips | -20°C, 6 months |
Reference Standard | 2 vials | |
Concentrated Biotinylated Detection Ab (100×) | 1 vial, 120 µL | |
Concentrated HRP Conjugate (100×) | 1 vial, 120 µL | -20°C(shading light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 4°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | |
HRP Conjugate Diluent | 1 vial, 14 mL | |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | |
Substrate Reagent A | 1 vial, 5 mL | 4°C (shading light) |
Substrate Reagent B | 1 vial, 5 mL | 4°C (shading light) |
Plate Sealer | 5 pieces | |
Product Description | 1 copy | |
Certificate of Analysis | 1 copy |
- Set standard, test sample and control (zero) wells on the pre-coated plate and record theirpositions. It is recommended to measure each standard and sample in duplicate. Note: addall solutions to the bottom of the plate wells while avoiding contact with the well walls. Ensuresolutions do not foam when adding to the wells.
- Aliquot 100µl of standard solutions into the standard wells.
- Add 100µl of Sample / Standard dilution buffer into the control (zero) well.
- Add 100µl of properly diluted sample (serum, plasma, tissue homogenates and otherbiological fluids. ) into test sample wells.
- Cover the plate with the sealer provided in the kit and incubate for 90 min at 37°C.
- Aspirate the liquid from each well, do not wash. Immediately add 100µL of BiotinylatedDetection Ab working solution to each well. Cover the plate with a plate seal and gently mix. Incubate for 1 hour at 37°C.
- Aspirate or decant the solution from the plate and add 350µL of wash buffer to each welland incubate for 1-2 minutes at room temperature. Aspirate the solution from each well andclap the plate on absorbent filter paper to dry. Repeat this process 3 times. Note: a microplatewasher can be used in this step and other wash steps.
- Add 100µL of HRP Conjugate working solution to each well. Cover with a plate seal andincubate for 30 min at 37°C.
- Aspirate or decant the solution from each well. Repeat the wash process for five times asconducted in step 7.
- Add 100µL of Substrate mixture solution to each well. Cover with a new plate seal andincubate for no more than 5 min at 37°C. Protect the plate from light.
- Determine the RLU value of each well immediately.