Mouse Cell Signalling ELISA Kits 2
Mouse IL-1ra CLIA Kit (MOES00035)
- SKU:
- MOES00035
- Product Type:
- ELISA Kit
- ELISA Type:
- CLIA Kit
- Size:
- 96 Assays
- Sensitivity:
- 7.5pg/mL
- Range:
- 12.5-800pg/mL
- ELISA Type:
- Sandwich
- Synonyms:
- IL1RN, DIRA, ICIL-1RA, IL-1RN, IL-1RA, IL-1ra3, IL1F3, IL1RA, IRAP, MVCD4
- Reactivity:
- Mouse
- Sample Type:
- Serum, plasma and other biological fluids
Description
Assay type: | Sandwich |
Format: | 96T |
Assay time: | 4.5h |
Reactivity: | Mouse |
Detection method: | Chemiluminescence |
Detection range: | 12.50-800 pg/mL |
Sensitivity: | 7.50 pg/mL |
Sample volume: | 100µL |
Sample type: | Serum, plasma and other biological fluids |
Repeatability: | CV < 15% |
Specificity: | This kit recognizes Mouse IL1ra CLIA Kit in samples. No significant cross-reactivity or interference between Mouse IL1ra CLIA Kit and analogues was observed. |
This kit uses Sandwich-CLIA as the method. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Mouse IL1ra. Standards or samples are added to the appropriate micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Mouse IL1ra and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well successively and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse IL1ra, biotinylated detection antibody and Avidin-HRP conjugate will appear fluorescence. The Relative light unit (RLU) value is measured spectrophotometrically by the Chemiluminescence immunoassay analyzer. The RLU value is positively associated with the concentration of Mouse IL1ra. The concentration of Mouse IL1ra in the samples can be calculated by comparing the RLU of the samples to the standard curve.
UniProt Protein Function: | IL1RN: Inhibits the activity of interleukin-1 by binding to receptor IL1R1 and preventing its association with the coreceptor IL1RAP for signaling. Has no interleukin-1 like activity. Binds functional interleukin-1 receptor IL1R1 with greater affinity than decoy receptor IL1R2; however, the physiological relevance of the latter association is unsure. The intracellular form of IL1RN is predominantly expressed in epithelial cells. 4 isoforms of the human protein are produced by alternative splicing. |
UniProt Protein Details: | Protein type:Secreted; Vesicle; Cytokine; Secreted, signal peptide Cellular Component: extracellular space; cell; cytoplasm; extracellular region; vesicle Molecular Function:interleukin-1 Type I receptor antagonist activity; interleukin-1 receptor binding; interleukin-1 Type II receptor antagonist activity; interleukin-1, Type II receptor binding; cytokine activity; interleukin-1, Type I receptor binding; interleukin-1 receptor antagonist activity Biological Process: positive regulation of JNK activity; negative regulation of glutamate secretion; negative regulation of heterotypic cell-cell adhesion; cytokine and chemokine mediated signaling pathway; response to glucocorticoid stimulus; sensory perception of pain; fever; memory; insulin secretion; acute-phase response; negative regulation of cytokine and chemokine mediated signaling pathway; lipid metabolic process; negative regulation of membrane potential; negative regulation of cell migration; negative regulation of apoptosis |
UniProt Code: | P25085 |
NCBI GenInfo Identifier: | 124313 |
NCBI Gene ID: | 16181 |
NCBI Accession: | P25085. 1 |
UniProt Secondary Accession: | P25085,O70207, |
UniProt Related Accession: | P25085 |
Molecular Weight: | 17,996 Da |
NCBI Full Name: | Interleukin-1 receptor antagonist protein |
NCBI Synonym Full Names: | interleukin 1 receptor antagonist |
NCBI Official Symbol: | Il1rn |
NCBI Official Synonym Symbols: | IL-1ra; F630041P17Rik |
NCBI Protein Information: | interleukin-1 receptor antagonist protein; IRAP; IL-1RN; IL1 inhibitor |
UniProt Protein Name: | Interleukin-1 receptor antagonist protein |
UniProt Synonym Protein Names: | IL1 inhibitor |
Protein Family: | Interleukin-1 receptor antagonist protein |
UniProt Gene Name: | Il1rn |
UniProt Entry Name: | IL1RA_MOUSE |
As the RLU values of the standard curve may vary according to the conditions of the actual assay performance (e. g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration (pg/mL) | RLU | Average | Corrected |
800 | 52595 52689 | 52642 | 52616 |
400 | 21502 21966 | 21734 | 21708 |
200 | 9923 9493 | 9708 | 9682 |
100 | 4210 4894 | 4552 | 4526 |
50 | 2271 2105 | 2188 | 2162 |
25 | 1090 1030 | 1060 | 1034 |
12.50 | 489 529 | 509 | 483 |
0 | 25 27 | 26 | -- |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Mouse IL1ra CLIA Kit were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Mouse IL1ra CLIA Kit were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 36.61 | 118.85 | 395.82 | 39.50 | 128.09 | 389.86 |
Standard deviation | 4.09 | 9.48 | 26.01 | 4.00 | 9.48 | 28.34 |
C V (%) | 11.17 | 7.98 | 6.57 | 10.13 | 7.40 | 7.27 |
Recovery
The recovery of Mouse IL1ra CLIA Kit spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
Serum (n=5) | 100-112 | 106 |
EDTA plasma (n=5) | 95-111 | 103 |
Cell culture media (n=5) | 98-110 | 104 |
Linearity
Samples were spiked with high concentrations of Mouse IL1ra CLIA Kit and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
1:2 | Range (%) | 92-105 | 88-99 | 100-116 |
Average (%) | 99 | 93 | 107 | |
1:4 | Range (%) | 86-101 | 94-109 | 84-99 |
Average (%) | 93 | 101 | 90 | |
1:8 | Range (%) | 101-115 | 96-108 | 92-105 |
Average (%) | 108 | 102 | 98 | |
1:16 | Range (%) | 99-113 | 95-110 | 101-113 |
Average (%) | 105 | 103 | 107 |
An unopened kit can be stored at 4°C for 1 month. If the kit is not used within 1 month, store the items separately according to the following conditions once the kit is received.
Item | Specifications | Storage |
Micro CLIA Plate(Dismountable) | 8 wells ×12 strips | -20°C, 6 months |
Reference Standard | 2 vials | |
Concentrated Biotinylated Detection Ab (100×) | 1 vial, 120 µL | |
Concentrated HRP Conjugate (100×) | 1 vial, 120 µL | -20°C(shading light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 4°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | |
HRP Conjugate Diluent | 1 vial, 14 mL | |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | |
Substrate Reagent A | 1 vial, 5 mL | 4°C (shading light) |
Substrate Reagent B | 1 vial, 5 mL | 4°C (shading light) |
Plate Sealer | 5 pieces | |
Product Description | 1 copy | |
Certificate of Analysis | 1 copy |
- Set standard, test sample and control (zero) wells on the pre-coated plate and record theirpositions. It is recommended to measure each standard and sample in duplicate. Note: addall solutions to the bottom of the plate wells while avoiding contact with the well walls. Ensuresolutions do not foam when adding to the wells.
- Aliquot 100 µL of standard solutions into the standard wells.
- Add 100 µL of Sample / Standard dilution buffer into the control (zero) well.
- Add 100 µL of properly diluted sample (serum, plasma, tissue homogenates and otherbiological fluids. ) into test sample wells.
- Cover the plate with the sealer provided in the kit and incubate for 90 min at 37 °C.
- Aspirate the liquid from each well, do not wash. Immediately add 100 µL of BiotinylatedDetection Ab working solution to each well. Cover the plate with a plate seal and gently mix. Incubate for 1 hour at 37 °C.
- Aspirate or decant the solution from the plate and add 350 µL of wash buffer to each welland incubate for 1-2 minutes at room temperature. Aspirate the solution from each well andclap the plate on absorbent filter paper to dry. Repeat this process 3 times. Note: a microplatewasher can be used in this step and other wash steps.
- Add 100 µL of HRP Conjugate working solution to each well. Cover with a plate seal andincubate for 30 min at 37 °C.
- Aspirate or decant the solution from each well. Repeat the wash process for five times asconducted in step 7.
- Add 100 µL of Substrate mixture solution to each well. Cover with a new plate seal andincubate for no more than 5 min at 37 °C. Protect the plate from light.
- Determine the RLU value of each well immediately.