Description
Assay type: | Sandwich |
Format: | 96T |
Assay time: | 4.5h |
Reactivity: | Mouse |
Detection method: | Chemiluminescence |
Detection range: | 62.50-4000 pg/mL |
Sensitivity: | 37.50 pg/mL |
Sample volume: | 100µL |
Sample type: | Serum, plasma and other biological fluids |
Repeatability: | CV < 15% |
Specificity: | This kit recognizes Mouse ICAM-1 in samples. No significant cross-reactivity or interference between Mouse ICAM-1 and analogues was observed. |
This kit uses Sandwich-CLIA as the method. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Mouse ICAM-1/CD54. Standards or samples are added to the appropriate micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Mouse ICAM-1/CD54 and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well successively and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Mouse ICAM-1/CD54, biotinylated detection antibody and Avidin-HRP conjugate will appear fluorescence. The Relative light unit (RLU) value is measured spectrophotometrically by the Chemiluminescence immunoassay analyzer. The RLU value is positively associated with the concentration of Mouse ICAM-1/CD54. The concentration of Mouse ICAM-1/CD54 in the samples can be calculated by comparing the RLU of the samples to the standard curve.
UniProt Protein Function: | ICAM1: a type I membrane protein of the immunoglobulin superfamily. Is a ligand for the leukocyte adhesion LFA-1 protein (Integrin alpha-L/beta-2) and a Rhinovirus receptor. Typically expressed on endothelial cells and cells of the immune system. ICAM1 binds to integrins of type CD11a / CD18, or CD11b / CD18. Its expression is activated by p53 in an NF-kappaB-independent manner. Induced by TNFalpha in a process that involves IKKbeta. |
UniProt Protein Details: | Protein type:Cell adhesion; Membrane protein, integral; Immunoglobulin superfamily Cellular Component: extracellular space; cell surface; focal adhesion; membrane; integral to plasma membrane; integral to membrane; plasma membrane; immunological synapse; external side of plasma membrane; lipid raft Molecular Function:integrin binding; protein complex binding Biological Process: regulation of cell adhesion; virion attachment, binding of host cell surface receptor; negative regulation of calcium ion transport; positive regulation of nitric oxide biosynthetic process; T cell antigen processing and presentation; activation of NF-kappaB transcription factor; positive regulation of cellular extravasation; regulation of cell shape; cell-cell adhesion; leukocyte adhesion; positive regulation of peptidyl-tyrosine phosphorylation; sensory perception of sound; positive regulation of actin filament polymerization; T cell activation via T cell receptor contact with antigen bound to MHC molecule on antigen presenting cell; positive regulation of vasoconstriction; adhesion to host; cell adhesion mediated by integrin; cell adhesion; acute inflammatory response to antigenic stimulus |
NCBI Summary: | This gene encodes an integral membrane protein that binds leukocyte adhesion protein LFA-1. It participates in the innate immune response. [provided by RefSeq, Aug 2016] |
UniProt Code: | P13597 |
NCBI GenInfo Identifier: | 124099 |
NCBI Gene ID: | 15894 |
NCBI Accession: | P13597. 1 |
UniProt Related Accession: | P13597 |
Molecular Weight: | |
NCBI Full Name: | Intercellular adhesion molecule 1 |
NCBI Synonym Full Names: | intercellular adhesion molecule 1 |
NCBI Official Symbol: | Icam1 |
NCBI Official Synonym Symbols: | CD54; Ly-47; Icam-1; MALA-2 |
NCBI Protein Information: | intercellular adhesion molecule 1 |
UniProt Protein Name: | Intercellular adhesion molecule 1 |
UniProt Synonym Protein Names: | MALA-2; MyD10; CD_antigen: CD54 |
UniProt Gene Name: | Icam1 |
UniProt Entry Name: | ICAM1_MOUSE |
As the RLU values of the standard curve may vary according to the conditions of the actual assay performance (e. g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration (pg/mL) | RLU | Average | Corrected |
4000 | 57792 60370 | 59081 | 59046 |
2000 | 26633 29085 | 27859 | 27824 |
1000 | 13833 13135 | 13484 | 13449 |
500 | 6382 6828 | 6605 | 6570 |
250 | 3494 2992 | 3243 | 3208 |
125 | 1622 1540 | 1581 | 1546 |
62.50 | 719 791 | 755 | 720 |
0 | 35 35 | 35 | -- |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Mouse ICAM-1 were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Mouse ICAM-1 were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 191.99 | 578.30 | 1886.02 | 207.53 | 546.87 | 1712.80 |
Standard deviation | 16.28 | 42.39 | 146.54 | 24.76 | 57.31 | 105.51 |
C V (%) | 8.48 | 7.33 | 7.77 | 11.93 | 10.48 | 6.16 |
Recovery
The recovery of Mouse ICAM-1 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
Serum (n=5) | 94-106 | 99 |
EDTA plasma (n=5) | 95-108 | 103 |
Cell culture media (n=5) | 92-105 | 98 |
Linearity
Samples were spiked with high concentrations of Mouse ICAM-1 and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
1:2 | Range (%) | 101-118 | 86-98 | 95-108 |
Average (%) | 107 | 91 | 101 | |
1:4 | Range (%) | 103-118 | 99-111 | 90-104 |
Average (%) | 109 | 104 | 95 | |
1:8 | Range (%) | 99-111 | 99-114 | 86-98 |
Average (%) | 106 | 106 | 91 | |
1:16 | Range (%) | 89-103 | 98-112 | 89-101 |
Average (%) | 95 | 106 | 94 |
An unopened kit can be stored at 4°C for 1 month. If the kit is not used within 1 month, store the items separately according to the following conditions once the kit is received.
Item | Specifications | Storage |
Micro CLIA Plate(Dismountable) | 8 wells ×12 strips | -20°C, 6 months |
Reference Standard | 2 vials | |
Concentrated Biotinylated Detection Ab (100×) | 1 vial, 120 µL | |
Concentrated HRP Conjugate (100×) | 1 vial, 120 µL | -20°C(shading light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 4°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | |
HRP Conjugate Diluent | 1 vial, 14 mL | |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | |
Substrate Reagent A | 1 vial, 5 mL | 4°C (shading light) |
Substrate Reagent B | 1 vial, 5 mL | 4°C (shading light) |
Plate Sealer | 5 pieces | |
Product Description | 1 copy | |
Certificate of Analysis | 1 copy |
- Set standard, test sample and control (zero) wells on the pre-coated plate and record theirpositions. It is recommended to measure each standard and sample in duplicate. Note: addall solutions to the bottom of the plate wells while avoiding contact with the well walls. Ensuresolutions do not foam when adding to the wells.
- Aliquot 100 µL of standard solutions into the standard wells.
- Add 100 µL of Sample / Standard dilution buffer into the control (zero) well.
- Add 100 µL of properly diluted sample (serum, plasma, tissue homogenates and otherbiological fluids. ) into test sample wells.
- Cover the plate with the sealer provided in the kit and incubate for 90 min at 37 °C.
- Aspirate the liquid from each well, do not wash. Immediately add 100 µL of BiotinylatedDetection Ab working solution to each well. Cover the plate with a plate seal and gently mix. Incubate for 1 hour at 37 °C.
- Aspirate or decant the solution from the plate and add 350 µL of wash buffer to each welland incubate for 1-2 minutes at room temperature. Aspirate the solution from each well andclap the plate on absorbent filter paper to dry. Repeat this process 3 times. Note: a microplatewasher can be used in this step and other wash steps.
- Add 100 µL of HRP Conjugate working solution to each well. Cover with a plate seal andincubate for 30 min at 37 °C.
- Aspirate or decant the solution from each well. Repeat the wash process for five times asconducted in step 7.
- Add 100 µL of Substrate mixture solution to each well. Cover with a new plate seal andincubate for no more than 5 min at 37 °C. Protect the plate from light.
- Determine the RLU value of each well immediately.