Human Epigenetics and Nuclear Signaling ELISA Kits
Human TK1 (Thymidine Kinase 1, Soluble) CLIA Kit (HUES01276)
- SKU:
- HUES01276
- Product Type:
- ELISA Kit
- ELISA Type:
- CLIA Kit
- Size:
- 96 Assays
- Sensitivity:
- 9.38pg/mL
- Range:
- 15.63-1000pg/mL
- ELISA Type:
- Sandwich
- Synonyms:
- TK1
- Reactivity:
- Human
- Sample Type:
- Serum, plasma and other biological fluids
- Research Area:
- Epigenetics and Nuclear Signaling
Description
Assay type: | Sandwich |
Format: | 96T |
Assay time: | 4.5h |
Reactivity: | Human |
Detection method: | Chemiluminescence |
Detection range: | 15.63-1000 pg/mL |
Sensitivity: | 9.38 pg/mL |
Sample volume: | 100µL |
Sample type: | Serum, plasma and other biological fluids |
Repeatability: | CV < 15% |
Specificity: | This kit recognizes Human TK1 in samples. No significant cross-reactivity or interference between Human TK1 and analogues was observed. |
This kit uses Sandwich-CLIA as the method. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human TK1. Standards or samples are added to the appropriate micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human TK1 and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well successively and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human TK1, biotinylated detection antibody and Avidin-HRP conjugate will appear fluorescence. The Relative light unit (RLU) value is measured spectrophotometrically by the Chemiluminescence immunoassay analyzer. The RLU value is positively associated with the concentration of Human TK1. The concentration of Human TK1 in the samples can be calculated by comparing the RLU of the samples to the standard curve.
UniProt Protein Function: | TK: cytosolic thymidine kinase. Phosphorylated during mitosis. Its enzymatic activity is high in proliferating cells and peaks during the S-phase of the cell cycle; it is very low in resting cells. |
UniProt Protein Details: | Protein type:EC 2. 7. 1. 21; Xenobiotic Metabolism - drug metabolism - other enzymes; Nucleotide Metabolism - pyrimidine; Kinase, other Chromosomal Location of Human Ortholog: 17q23. 2-q25. 3 Cellular Component: cytosol Molecular Function:identical protein binding; nucleoside kinase activity; zinc ion binding; thymidine kinase activity; ATP binding Biological Process: response to nutrient levels; skeletal muscle cell proliferation; nucleobase, nucleoside and nucleotide metabolic process; response to toxin; liver development; protein homotetramerization; fetal process involved in parturition; deoxyribonucleoside monophosphate biosynthetic process; response to cortisol stimulus; pyrimidine base metabolic process; gut development; response to copper ion; nucleotide biosynthetic process; nucleobase, nucleoside, nucleotide and nucleic acid metabolic process; thymidine metabolic process; pyrimidine nucleoside salvage; phosphorylation |
UniProt Code: | P04183 |
NCBI GenInfo Identifier: | 23503074 |
NCBI Gene ID: | 7083 |
NCBI Accession: | P04183. 2 |
UniProt Secondary Accession: | P04183,Q969V0, Q9UMG9, B2RC58, |
UniProt Related Accession: | P04183 |
Molecular Weight: | 234 |
NCBI Full Name: | Thymidine kinase, cytosolic |
NCBI Synonym Full Names: | thymidine kinase 1, soluble |
NCBI Official Symbol: | TK1 |
NCBI Official Synonym Symbols: | TK2 |
NCBI Protein Information: | thymidine kinase, cytosolic; thymidine kinase-1; thymidine kinase 1 soluble isoform |
UniProt Protein Name: | Thymidine kinase, cytosolic |
Protein Family: | Thymidine kinase |
UniProt Gene Name: | TK1 |
UniProt Entry Name: | KITH_HUMAN |
As the RLU values of the standard curve may vary according to the conditions of the actual assay performance (e. g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration (pg/mL) | RLU | Average | Corrected |
1000 | 52562 54788 | 53675 | 53650 |
500 | 22338 22640 | 22489 | 22464 |
250 | 10735 9687 | 10211 | 10186 |
125 | 4759 5043 | 4901 | 4876 |
62.5 | 2547 2359 | 2453 | 2428 |
31.25 | 1300 1262 | 1281 | 1256 |
15.63 | 692 724 | 708 | 683 |
0 | 24 26 | 25 | -- |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Human TK1 were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Human TK1 were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 51.81 | 133.49 | 445.01 | 54.75 | 141.81 | 408.20 |
Standard deviation | 5.90 | 13.40 | 32.17 | 6.34 | 10.95 | 45.84 |
C V (%) | 11.39 | 10.04 | 7.23 | 11.58 | 7.72 | 11.23 |
Recovery
The recovery of Human TK1 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
Serum (n=5) | 90-105 | 97 |
EDTA plasma (n=5) | 99-113 | 106 |
Cell culture media (n=5) | 98-112 | 106 |
Linearity
Samples were spiked with high concentrations of Human TK1 and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
1:2 | Range (%) | 99-113 | 99-110 | 85-97 |
Average (%) | 107 | 104 | 91 | |
1:4 | Range (%) | 97-111 | 89-100 | 100-117 |
Average (%) | 105 | 94 | 108 | |
1:8 | Range (%) | 97-110 | 93-106 | 91-106 |
Average (%) | 102 | 99 | 97 | |
1:16 | Range (%) | 99-115 | 86-96 | 87-103 |
Average (%) | 105 | 91 | 94 |
An unopened kit can be stored at 4°C for 1 month. If the kit is not used within 1 month, store the items separately according to the following conditions once the kit is received.
Item | Specifications | Storage |
Micro CLIA Plate(Dismountable) | 8 wells ×12 strips | -20°C, 6 months |
Reference Standard | 2 vials | |
Concentrated Biotinylated Detection Ab (100×) | 1 vial, 120 µL | |
Concentrated HRP Conjugate (100×) | 1 vial, 120 µL | -20°C(shading light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 4°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | |
HRP Conjugate Diluent | 1 vial, 14 mL | |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | |
Substrate Reagent A | 1 vial, 5 mL | 4°C (shading light) |
Substrate Reagent B | 1 vial, 5 mL | 4°C (shading light) |
Plate Sealer | 5 pieces | |
Product Description | 1 copy | |
Certificate of Analysis | 1 copy |
- Set standard, test sample and control (zero) wells on the pre-coated plate and record theirpositions. It is recommended to measure each standard and sample in duplicate. Note: addall solutions to the bottom of the plate wells while avoiding contact with the well walls. Ensuresolutions do not foam when adding to the wells.
- Aliquot 100 µL of standard solutions into the standard wells.
- Add 100 µL of Sample / Standard dilution buffer into the control (zero) well.
- Add 100 µL of properly diluted sample (serum, plasma, tissue homogenates and otherbiological fluids. ) into test sample wells.
- Cover the plate with the sealer provided in the kit and incubate for 90 min at 37 °C.
- Aspirate the liquid from each well, do not wash. Immediately add 100 µL of BiotinylatedDetection Ab working solution to each well. Cover the plate with a plate seal and gently mix. Incubate for 1 hour at 37 °C.
- Aspirate or decant the solution from the plate and add 350 µL of wash buffer to each welland incubate for 1-2 minutes at room temperature. Aspirate the solution from each well andclap the plate on absorbent filter paper to dry. Repeat this process 3 times. Note: a microplatewasher can be used in this step and other wash steps.
- Add 100 µL of HRP Conjugate working solution to each well. Cover with a plate seal andincubate for 30 min at 37 °C.
- Aspirate or decant the solution from each well. Repeat the wash process for five times asconducted in step 7.
- Add 100 µL of Substrate mixture solution to each well. Cover with a new plate seal andincubate for no more than 5 min at 37 °C. Protect the plate from light.
- Determine the RLU value of each well immediately.