Human Developmental Biology ELISA Kits
Human STMN1 (Stathmin 1) CLIA Kit (HUES01093)
- SKU:
- HUES01093
- Product Type:
- ELISA Kit
- ELISA Type:
- CLIA Kit
- Size:
- 96 Assays
- Sensitivity:
- 18.75pg/mL
- Range:
- 31.25-2000pg/mL
- ELISA Type:
- Sandwich
- Reactivity:
- Human
- Sample Type:
- Serum, plasma and other biological fluids
- Research Area:
- Developmental Biology
Description
Assay type: | Sandwich |
Format: | 96T |
Assay time: | 4.5h |
Reactivity: | Human |
Detection method: | Chemiluminescence |
Detection range: | 31.25-2000 pg/mL |
Sensitivity: | 18.75 pg/mL |
Sample volume: | 100µL |
Sample type: | Serum, plasma and other biological fluids |
Repeatability: | CV < 15% |
Specificity: | This kit recognizes Human STMN1 in samples. No significant cross-reactivity or interference between Human STMN1 and analogues was observed. |
This kit uses Sandwich-CLIA as the method. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human STMN1. Standards or samples are added to the appropriate micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human STMN1 and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well successively and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human STMN1, biotinylated detection antibody and Avidin-HRP conjugate will appear fluorescence. The Relative light unit (RLU) value is measured spectrophotometrically by the Chemiluminescence immunoassay analyzer. The RLU value is positively associated with the concentration of Human STMN1. The concentration of Human STMN1 in the samples can be calculated by comparing the RLU of the samples to the standard curve.
UniProt Protein Function: | STMN1: a microtubule-associated protein involved in the regulation of the microtubule filament system by destabilizing microtubules. It prevents assembly and promotes disassembly of microtubules. Phosphorylation reduces tubulin binding 10-fold and suppresses the MT polymerization inhibition activity. Present in much greater abundance in cells from patients with acute leukemia of different subtypes than in normal peripheral blood lymphocytes, nonleukemic proliferating lymphoid cells, bone marrow cells, or cells from patients with chronic lymphoid or myeloid leukemia. |
UniProt Protein Details: | Protein type:Cytoskeletal Chromosomal Location of Human Ortholog: 1p36. 11 Cellular Component: cytoplasm; cytosol; intracellular; membrane; microtubule; neuron projection Molecular Function:protein binding; signal transducer activity; tubulin binding Biological Process: axonogenesis; brain development; microtubule depolymerization; mitotic spindle organization and biogenesis; negative regulation of microtubule polymerization; neurite development; positive regulation of cell motility; regulation of cytoskeleton organization and biogenesis; response to virus; signal transduction |
NCBI Summary: | This gene belongs to the stathmin family of genes. It encodes a ubiquitous cytosolic phosphoprotein proposed to function as an intracellular relay integrating regulatory signals of the cellular environment. The encoded protein is involved in the regulation of the microtubule filament system by destabilizing microtubules. It prevents assembly and promotes disassembly of microtubules. Multiple transcript variants encoding different isoforms have been found for this gene. [provided by RefSeq, Feb 2009] |
UniProt Code: | P16949 |
NCBI GenInfo Identifier: | 134973 |
NCBI Gene ID: | 3925 |
NCBI Accession: | P16949. 3 |
UniProt Secondary Accession: | P16949,A2A2D1, B2R4E7, B7Z8N4, D3DPJ5, |
UniProt Related Accession: | P16949 |
Molecular Weight: | 19,824 Da |
NCBI Full Name: | Stathmin |
NCBI Synonym Full Names: | stathmin 1 |
NCBI Official Symbol: | STMN1 |
NCBI Official Synonym Symbols: | Lag; SMN; OP18; PP17; PP19; PR22; LAP18; C1orf215 |
NCBI Protein Information: | stathmin |
UniProt Protein Name: | Stathmin |
UniProt Synonym Protein Names: | Leukemia-associated phosphoprotein p18; Metablastin; Oncoprotein 18; Op18; Phosphoprotein p19; pp19; Prosolin; Protein Pr22; pp17 |
Protein Family: | Stathmin |
UniProt Gene Name: | STMN1 |
UniProt Entry Name: | STMN1_HUMAN |
As the RLU values of the standard curve may vary according to the conditions of the actual assay performance (e. g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration (pg/mL) | RLU | Average | Corrected |
2000 | 54521 56921 | 55721 | 55693 |
1000 | 23295 24653 | 23974 | 23946 |
500 | 11134 10930 | 11032 | 11004 |
250 | 5029 5559 | 5294 | 5266 |
125 | 2622 2594 | 2608 | 2580 |
62.5 | 1384 1238 | 1311 | 1283 |
31.25 | 665 683 | 674 | 646 |
0 | 27 29 | 28 | -- |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Human STMN1 were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Human STMN1 were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 95.30 | 203.88 | 912.83 | 89.42 | 201.39 | 904.63 |
Standard deviation | 11.86 | 19.31 | 74.94 | 7.28 | 17.54 | 86.48 |
C V (%) | 12.44 | 9.47 | 8.21 | 8.14 | 8.71 | 9.56 |
Recovery
The recovery of Human STMN1 spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
Serum (n=5) | 99-115 | 107 |
EDTA plasma (n=5) | 91-104 | 97 |
Cell culture media (n=5) | 94-108 | 99 |
Linearity
Samples were spiked with high concentrations of Human STMN1 and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
1:2 | Range (%) | 97-109 | 101-119 | 101-116 |
Average (%) | 102 | 108 | 107 | |
1:4 | Range (%) | 94-107 | 100-112 | 87-101 |
Average (%) | 101 | 107 | 94 | |
1:8 | Range (%) | 97-110 | 85-98 | 98-116 |
Average (%) | 104 | 92 | 106 | |
1:16 | Range (%) | 97-113 | 89-101 | 90-101 |
Average (%) | 105 | 94 | 95 |
An unopened kit can be stored at 4°C for 1 month. If the kit is not used within 1 month, store the items separately according to the following conditions once the kit is received.
Item | Specifications | Storage |
Micro CLIA Plate(Dismountable) | 8 wells ×12 strips | -20°C, 6 months |
Reference Standard | 2 vials | |
Concentrated Biotinylated Detection Ab (100×) | 1 vial, 120 µL | |
Concentrated HRP Conjugate (100×) | 1 vial, 120 µL | -20°C(shading light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 4°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | |
HRP Conjugate Diluent | 1 vial, 14 mL | |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | |
Substrate Reagent A | 1 vial, 5 mL | 4°C (shading light) |
Substrate Reagent B | 1 vial, 5 mL | 4°C (shading light) |
Plate Sealer | 5 pieces | |
Product Description | 1 copy | |
Certificate of Analysis | 1 copy |
- Set standard, test sample and control (zero) wells on the pre-coated plate and record theirpositions. It is recommended to measure each standard and sample in duplicate. Note: addall solutions to the bottom of the plate wells while avoiding contact with the well walls. Ensuresolutions do not foam when adding to the wells.
- Aliquot 100 µL of standard solutions into the standard wells.
- Add 100 µL of Sample / Standard dilution buffer into the control (zero) well.
- Add 100 µL of properly diluted sample (serum, plasma, tissue homogenates and otherbiological fluids. ) into test sample wells.
- Cover the plate with the sealer provided in the kit and incubate for 90 min at 37 °C.
- Aspirate the liquid from each well, do not wash. Immediately add 100 µL of BiotinylatedDetection Ab working solution to each well. Cover the plate with a plate seal and gently mix. Incubate for 1 hour at 37 °C.
- Aspirate or decant the solution from the plate and add 350 µL of wash buffer to each welland incubate for 1-2 minutes at room temperature. Aspirate the solution from each well andclap the plate on absorbent filter paper to dry. Repeat this process 3 times. Note: a microplatewasher can be used in this step and other wash steps.
- Add 100 µL of HRP Conjugate working solution to each well. Cover with a plate seal andincubate for 30 min at 37 °C.
- Aspirate or decant the solution from each well. Repeat the wash process for five times asconducted in step 7.
- Add 100 µL of Substrate mixture solution to each well. Cover with a new plate seal andincubate for no more than 5 min at 37 °C. Protect the plate from light.
- Determine the RLU value of each well immediately.