Human Immunology ELISA Kits 1
Human PGRP-S (Peptidoglycan Recognition Protein-S) CLIA Kit (HUES01077)
- SKU:
- HUES01077
- Product Type:
- ELISA Kit
- ELISA Type:
- CLIA Kit
- Size:
- 96 Assays
- Sensitivity:
- 37.5pg/mL
- Range:
- 62.5-4000pg/mL
- ELISA Type:
- Sandwich
- Reactivity:
- Human
- Sample Type:
- Serum, plasma and other biological fluids
- Research Area:
- Immunology
Description
Assay type: | Sandwich |
Format: | 96T |
Assay time: | 4.5h |
Reactivity: | Human |
Detection method: | Chemiluminescence |
Detection range: | 62.50-4000 pg/mL |
Sensitivity: | 37.50 pg/mL |
Sample volume: | 100µL |
Sample type: | Serum, plasma and other biological fluids |
Repeatability: | CV < 15% |
Specificity: | This kit recognizes Human PGRP-S in samples. No significant cross-reactivity or interference between Human PGRP-S and analogues was observed. |
This kit uses Sandwich-CLIA as the method. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human PGRP-S. Standards or samples are added to the appropriate micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human PGRP-S and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well successively and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human PGRP-S, biotinylated detection antibody and Avidin-HRP conjugate will appear fluorescence. The Relative light unit (RLU) value is measured spectrophotometrically by the Chemiluminescence immunoassay analyzer. The RLU value is positively associated with the concentration of Human PGRP-S. The concentration of Human PGRP-S in the samples can be calculated by comparing the RLU of the samples to the standard curve.
UniProt Protein Function: | PGLYRP1: Pattern receptor that binds to murein peptidoglycans (PGN) of Gram-positive bacteria. Has bactericidal activity towards Gram-positive bacteria. May kill Gram-positive bacteria by interfering with peptidoglycan biosynthesis. Binds also to Gram- negative bacteria, and has bacteriostatic activity towards Gram- negative bacteria. Plays a role in innate immunity. Belongs to the N-acetylmuramoyl-L-alanine amidase 2 family. |
UniProt Protein Details: | Protein type:Secreted, signal peptide; Apoptosis; Receptor, misc. ; Secreted Chromosomal Location of Human Ortholog: 19q13. 2-q13. 3 Cellular Component: extracellular region Molecular Function:peptidoglycan receptor activity; peptidoglycan binding; zinc ion binding; N-acetylmuramoyl-L-alanine amidase activity Biological Process: defense response to Gram-positive bacterium; negative regulation of natural killer cell differentiation during immune response; detection of bacterium; negative regulation of interferon-gamma production; negative regulation of inflammatory response; innate immune response; immune response; pattern recognition receptor signaling pathway; peptidoglycan catabolic process |
UniProt Code: | O75594 |
NCBI GenInfo Identifier: | 18202143 |
NCBI Gene ID: | 8993 |
NCBI Accession: | O75594. 1 |
UniProt Secondary Accession: | O75594,Q4VB36, |
UniProt Related Accession: | O75594 |
Molecular Weight: | |
NCBI Full Name: | Peptidoglycan recognition protein 1 |
NCBI Synonym Full Names: | peptidoglycan recognition protein 1 |
NCBI Official Symbol: | PGLYRP1 |
NCBI Official Synonym Symbols: | PGRP; TAG7; PGRPS; PGLYRP; PGRP-S; TNFSF3L |
NCBI Protein Information: | peptidoglycan recognition protein 1; TNF superfamily, member 3 (LTB)-like (peptidoglycan recognition protein) |
UniProt Protein Name: | Peptidoglycan recognition protein 1 |
UniProt Synonym Protein Names: | Peptidoglycan recognition protein short; PGRP-S |
Protein Family: | Peptidoglycan recognition protein |
UniProt Gene Name: | PGLYRP1 |
UniProt Entry Name: | PGRP1_HUMAN |
As the RLU values of the standard curve may vary according to the conditions of the actual assay performance (e. g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration (pg/mL) | RLU | Average | Corrected |
4000 | 54399 63763 | 59081 | 59046 |
2000 | 25244 30474 | 27859 | 27824 |
1000 | 13582 13386 | 13484 | 13449 |
500 | 6378 6832 | 6605 | 6570 |
250 | 3396 3090 | 3243 | 3208 |
125 | 1615 1547 | 1581 | 1546 |
62.50 | 752 758 | 755 | 720 |
0 | 35 35 | 35 | -- |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Human PGRP-S were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Human PGRP-S were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 184.68 | 573.70 | 1913.90 | 196.50 | 592.20 | 1824.75 |
Standard deviation | 17.12 | 63.68 | 180.86 | 19.22 | 52.00 | 132.66 |
C V (%) | 9.27 | 11.10 | 9.45 | 9.78 | 8.78 | 7.27 |
Recovery
The recovery of Human PGRP-S spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
Serum (n=5) | 90-106 | 96 |
EDTA plasma (n=5) | 96-112 | 102 |
Cell culture media (n=5) | 86-99 | 92 |
Linearity
Samples were spiked with high concentrations of Human PGRP-S and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
1:2 | Range (%) | 101-115 | 90-102 | 100-115 |
Average (%) | 108 | 97 | 107 | |
1:4 | Range (%) | 95-108 | 91-104 | 98-115 |
Average (%) | 100 | 98 | 105 | |
1:8 | Range (%) | 91-108 | 99-112 | 98-114 |
Average (%) | 99 | 105 | 104 | |
1:16 | Range (%) | 102-118 | 96-112 | 104-115 |
Average (%) | 110 | 102 | 109 |
An unopened kit can be stored at 4°C for 1 month. If the kit is not used within 1 month, store the items separately according to the following conditions once the kit is received.
Item | Specifications | Storage |
Micro CLIA Plate(Dismountable) | 8 wells ×12 strips | -20°C, 6 months |
Reference Standard | 2 vials | |
Concentrated Biotinylated Detection Ab (100×) | 1 vial, 120 µL | |
Concentrated HRP Conjugate (100×) | 1 vial, 120 µL | -20°C(shading light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 4°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | |
HRP Conjugate Diluent | 1 vial, 14 mL | |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | |
Substrate Reagent A | 1 vial, 5 mL | 4°C (shading light) |
Substrate Reagent B | 1 vial, 5 mL | 4°C (shading light) |
Plate Sealer | 5 pieces | |
Product Description | 1 copy | |
Certificate of Analysis | 1 copy |
- Set standard, test sample and control (zero) wells on the pre-coated plate and record theirpositions. It is recommended to measure each standard and sample in duplicate. Note: addall solutions to the bottom of the plate wells while avoiding contact with the well walls. Ensuresolutions do not foam when adding to the wells.
- Aliquot 100 µL of standard solutions into the standard wells.
- Add 100 µL of Sample / Standard dilution buffer into the control (zero) well.
- Add 100 µL of properly diluted sample (serum, plasma, tissue homogenates and otherbiological fluids. ) into test sample wells.
- Cover the plate with the sealer provided in the kit and incubate for 90 min at 37 °C.
- Aspirate the liquid from each well, do not wash. Immediately add 100 µL of BiotinylatedDetection Ab working solution to each well. Cover the plate with a plate seal and gently mix. Incubate for 1 hour at 37 °C.
- Aspirate or decant the solution from the plate and add 350 µL of wash buffer to each welland incubate for 1-2 minutes at room temperature. Aspirate the solution from each well andclap the plate on absorbent filter paper to dry. Repeat this process 3 times. Note: a microplatewasher can be used in this step and other wash steps.
- Add 100 µL of HRP Conjugate working solution to each well. Cover with a plate seal andincubate for 30 min at 37 °C.
- Aspirate or decant the solution from each well. Repeat the wash process for five times asconducted in step 7.
- Add 100 µL of Substrate mixture solution to each well. Cover with a new plate seal andincubate for no more than 5 min at 37 °C. Protect the plate from light.
- Determine the RLU value of each well immediately.