Human Cell Biology ELISA Kits 5
Human MGP (Matrix Gla Protein) CLIA Kit (HUES00817)
- SKU:
- HUES00817
- Product Type:
- ELISA Kit
- ELISA Type:
- CLIA Kit
- Size:
- 96 Assays
- Sensitivity:
- 4.69pg/mL
- Range:
- 7.81-500pg/mL
- ELISA Type:
- Sandwich
- Reactivity:
- Human
- Sample Type:
- Serum, plasma and other biological fluids
- Research Area:
- Cell Biology
Description
Assay type: | Sandwich |
Format: | 96T |
Assay time: | 4.5h |
Reactivity: | Human |
Detection method: | Chemiluminescence |
Detection range: | 7.81-500 pg/mL |
Sensitivity: | 4.69 pg/mL |
Sample volume: | 100µL |
Sample type: | Serum, plasma and other biological fluids |
Repeatability: | CV < 15% |
Specificity: | This kit recognizes Human MGP in samples. No significant cross-reactivity or interference between Human MGP and analogues was observed. |
This kit uses Sandwich-CLIA as the method. The micro CLIA plate provided in this kit has been pre-coated with an antibody specific to Human MGP. Standards or samples are added to the appropriate micro CLIA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human MGP and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well successively and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human MGP, biotinylated detection antibody and Avidin-HRP conjugate will appear fluorescence. The Relative light unit (RLU) value is measured spectrophotometrically by the Chemiluminescence immunoassay analyzer. The RLU value is positively associated with the concentration of Human MGP. The concentration of Human MGP in the samples can be calculated by comparing the RLU of the samples to the standard curve.
UniProt Protein Function: | MGP: Associates with the organic matrix of bone and cartilage. Thought to act as an inhibitor of bone formation. Defects in MGP are the cause of Keutel syndrome (KS). KS is an autosomal recessive disorder characterized by abnormal cartilage calcification, peripheral pulmonary stenosis neural hearing loss and midfacial hypoplasia. Belongs to the osteocalcin/matrix Gla protein family. |
UniProt Protein Details: | Protein type:Secreted; Extracellular matrix; Secreted, signal peptide Chromosomal Location of Human Ortholog: 12p12. 3 Cellular Component: extracellular matrix; proteinaceous extracellular matrix Molecular Function:protein binding; structural constituent of bone; extracellular matrix structural constituent; calcium ion binding Biological Process: ossification; cell differentiation; regulation of bone mineralization; cartilage condensation Disease: Keutel Syndrome |
NCBI Summary: | The protein encoded by this gene is secreted and likely acts as an inhibitor of bone formation. The encoded protein is found in the organic matrix of bone and cartilage. Defects in this gene are a cause of Keutel syndrome (KS). Two transcript variants encoding different isoforms have been found for this gene. [provided by RefSeq, Jun 2010] |
UniProt Code: | P08493 |
NCBI GenInfo Identifier: | 118572668 |
NCBI Gene ID: | 4256 |
NCBI Accession: | P08493. 2 |
UniProt Secondary Accession: | P08493,Q2TU41, Q567P9, Q6ICN5, A0M8W5, B2R519, J3KMX7 |
UniProt Related Accession: | P08493 |
Molecular Weight: | 103 |
NCBI Full Name: | Matrix Gla protein |
NCBI Synonym Full Names: | matrix Gla protein |
NCBI Official Symbol: | MGP |
NCBI Official Synonym Symbols: | NTI; GIG36; MGLAP |
NCBI Protein Information: | matrix Gla protein; cell growth-inhibiting gene 36 protein |
UniProt Protein Name: | Matrix Gla protein |
UniProt Synonym Protein Names: | Cell growth-inhibiting gene 36 protein |
Protein Family: | Matrix Gla protein |
UniProt Gene Name: | MGP |
UniProt Entry Name: | MGP_HUMAN |
As the RLU values of the standard curve may vary according to the conditions of the actual assay performance (e. g. operator, pipetting technique, washing technique or temperature effects), the operator should establish a standard curve for each test. Typical standard curve and data is provided below for reference only.
Concentration (pg/mL) | RLU | Average | Corrected |
500 | 47395 55679 | 51537 | 51508 |
250 | 19319 22677 | 20998 | 20969 |
125 | 9444 9348 | 9396 | 9367 |
62.5 | 4368 4656 | 4512 | 4483 |
31.25 | 2344 2254 | 2299 | 2270 |
15.63 | 1262 1238 | 1250 | 1221 |
7.81 | 709 771 | 740 | 711 |
0 | 28 30 | 29 | -- |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Human MGP were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Human MGP were tested on 3 different plates, 20 replicates in each plate.
Intra-assay Precision | Inter-assay Precision | |||||
Sample | 1 | 2 | 3 | 1 | 2 | 3 |
n | 20 | 20 | 20 | 20 | 20 | 20 |
Mean (pg/mL) | 22.89 | 79.32 | 204.58 | 22.05 | 86.53 | 219.41 |
Standard deviation | 2.63 | 9.26 | 21.17 | 2.10 | 9.82 | 18.32 |
C V (%) | 11.49 | 11.67 | 10.35 | 9.52 | 11.35 | 8.35 |
Recovery
The recovery of Human MGP spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Sample Type | Range (%) | Average Recovery (%) |
Serum (n=5) | 85-96 | 91 |
EDTA plasma (n=5) | 88-101 | 94 |
Cell culture media (n=5) | 86-97 | 92 |
Linearity
Samples were spiked with high concentrations of Human MGP and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
1:2 | Range (%) | 102-118 | 89-101 | 90-106 |
Average (%) | 107 | 96 | 97 | |
1:4 | Range (%) | 98-112 | 92-106 | 93-106 |
Average (%) | 104 | 98 | 99 | |
1:8 | Range (%) | 95-110 | 99-117 | 96-110 |
Average (%) | 101 | 107 | 104 | |
1:16 | Range (%) | 95-110 | 97-110 | 99-115 |
Average (%) | 101 | 104 | 106 |
An unopened kit can be stored at 4°C for 1 month. If the kit is not used within 1 month, store the items separately according to the following conditions once the kit is received.
Item | Specifications | Storage |
Micro CLIA Plate(Dismountable) | 8 wells ×12 strips | -20°C, 6 months |
Reference Standard | 2 vials | |
Concentrated Biotinylated Detection Ab (100×) | 1 vial, 120 µL | |
Concentrated HRP Conjugate (100×) | 1 vial, 120 µL | -20°C(shading light), 6 months |
Reference Standard & Sample Diluent | 1 vial, 20 mL | 4°C, 6 months |
Biotinylated Detection Ab Diluent | 1 vial, 14 mL | |
HRP Conjugate Diluent | 1 vial, 14 mL | |
Concentrated Wash Buffer (25×) | 1 vial, 30 mL | |
Substrate Reagent A | 1 vial, 5 mL | 4°C (shading light) |
Substrate Reagent B | 1 vial, 5 mL | 4°C (shading light) |
Plate Sealer | 5 pieces | |
Product Description | 1 copy | |
Certificate of Analysis | 1 copy |
- Set standard, test sample and control (zero) wells on the pre-coated plate and record theirpositions. It is recommended to measure each standard and sample in duplicate. Note: addall solutions to the bottom of the plate wells while avoiding contact with the well walls. Ensuresolutions do not foam when adding to the wells.
- Aliquot 100µl of standard solutions into the standard wells.
- Add 100µl of Sample / Standard dilution buffer into the control (zero) well.
- Add 100µl of properly diluted sample (serum, plasma, tissue homogenates and otherbiological fluids. ) into test sample wells.
- Cover the plate with the sealer provided in the kit and incubate for 90 min at 37°C.
- Aspirate the liquid from each well, do not wash. Immediately add 100µL of BiotinylatedDetection Ab working solution to each well. Cover the plate with a plate seal and gently mix. Incubate for 1 hour at 37°C.
- Aspirate or decant the solution from the plate and add 350µL of wash buffer to each welland incubate for 1-2 minutes at room temperature. Aspirate the solution from each well andclap the plate on absorbent filter paper to dry. Repeat this process 3 times. Note: a microplatewasher can be used in this step and other wash steps.
- Add 100µL of HRP Conjugate working solution to each well. Cover with a plate seal andincubate for 30 min at 37°C.
- Aspirate or decant the solution from each well. Repeat the wash process for five times asconducted in step 7.
- Add 100µL of Substrate mixture solution to each well. Cover with a new plate seal andincubate for no more than 5 min at 37°C. Protect the plate from light.
- Determine the RLU value of each well immediately.