Food Science

ERY (Erythromycin) ELISA Kit (FSES0055)

(No reviews yet) Write a Review
SKU:
FSES0055
Product Type:
ELISA Kit
Applications:
ELISA
Research Area:
Food Science

Description

Technical Manual

ERY (Erythromycin) ELISA Kit

Kit Name: ERY (Erythromycin) ELISA Kit
Kit Size: 96 Assays
Test Principle: This kit uses Competitive-ELISA as the method. It can detect Erythromycin (ERY) in samples, such as muscle, serum, etc. This kit is composed of ELISA Microtiter plate, HRP conjugate, antibody working solution, standard and other supplementary reagents. The microtiter plate in this kit has been pre-coated with coupled antigen. During the reaction, ERY in the samples or standard competes with coupled antigen on the solid phase supporter for sites of anti-ERY antibody. Then Horseradish Peroxidase (HRP) conjugate is added to each microtiter plate well, and Substrate Reagent is added for color development. There is a negative correlation between the OD value of samples and the concentration of ERY. The concentration of ERY in the samples can be calculated by comparing the OD of the samples to the standard curve.
Sensitivity: 0.1 ppb (ng/mL)
Reaction Mode: 25°C, 30min~30min~15 min
Detection Limit: Raw milk, Eggs ---2 ppb; Urine, Serum, Muscle (livestock) ---5 ppb. Muscle (Fish, shrimp) --10 ppb.
Cross Reactivity: Erythromycin ----100% Spiramycin,Valnemulin,Lincomycin,Kanamycin,Tilmicosin,Doramectin,Ivermectin, Eprinomectin, Moxidectin, Abamectin, Tylosin-----<0.1%
Sample Recovery Rate: 90%±30%.
Kits Components:
Item Specifications
ELISA Microtiter Plate 96 wells
Standard Liquid 1 mL each<  (0 ppb, 0.1 ppb, 0.3 ppb, 0.9 ppb, 2.7 ppb, 8.1 ppb)
HRP Conjugate 12 mL
Antibody Working Solution 10 mL
Substrate Reagent A 6 mL
Substrate Reagent B 6 mL
Stop Solution 6 mL
20×Concentrated Wash Buffer 25 mL
Tissue Dilution 50 mL
Plate Sealer 3 pieces
Sealed Bag 1 piece
Manual 1 copy
Instruments: Microplate reader, Oscillators, Centrifuge, Balance (sensibility 0.01 g).
Micropipette: Single channel (20-200 uL, 100-1000 uL), Multichannel (30-300 uL).
Reagents: Methanol, N, N-Dimethylformamide (DMF), NaCl, NaOH, Concentrated HCl.
Storage And Expiry Date: Store the kit at 2~8°C. Do not freeze any test kit components. Return any unused microwells to their original foil bag and reseal them together with the desiccant provided and further store at 2 - 8°C. Expiry date: expiration date is on the packing box.
View AllClose